AOIP OM 16 MIKROOHMMETRE

Mikroohmmetre



Marka : AOIP
Model : OM 16
AOIP OM 16 MIKROOHMMETRE
Özellikler

Performances and technical specifications @23°C ±5°C

Uncertainty is given in % of reading + fixed value.

Resistance measurement

Measurement range Resolution Accuracy / 1 year
(23°C ±5°C)
Measuring current Voltage drop
5 mΩ 0.1 μΩ 0.05 % + 0.5 μΩ 10 A 50 mV
25 mΩ 1 μΩ 0.05 % + 3 μΩ 10 A 250 mV
250 mΩ 10 μΩ 0.05 % + 30 μΩ 10 A 2.5 V
2500 mΩ 0.1 mΩ 0.05 % + 0.3 mΩ 1 A 2.5 V
25 Ω 1 mΩ 0.05 % + 3 mΩ 100 mA 2.5 V
250 Ω 10 mΩ 0.05 % + 30 mΩ 10 mA 2.5 V
2500 Ω 100 mΩ 0.05 % + 300 mΩ 1 mA 2.5 V

Automatic or manual selection of measurement range
Possible excess over the nominal range:

  • 5 mΩ range: + 20 %
  • 25 mΩ range: + 20 %

Maximum voltage between the terminals in an open circuit: 7 V
Current waveform: Continuous or pulse DC current

Ambient temperature measurement for Tref compensation

Type Résolution Précision sur 1 an
(23°C ±5°C)
Remarque
Pt100 0,1°C 0,5°C Mesure par Pt100 externe ou valeur saisie au clavier

Further features

Resistance types
  • Inductive resistances: Coils, transformers, motor windings, twisted cables...
  • Non-inductive resistances: Earth bonding, coating, contact résistances, non-twisted cables...
Measurement trigger conditions Manual or automatic trigger allowing a single operator to be able to perform measurements
EMFs Automatic compensation of EMF parasites before each measurement for a greater accuracy
Temperature compensation - Ambient temperature Tamb, measured with external Pt100 (or entered by the user
- Programmed reference temperature Tref, to which the measured value is converted:
R(Tref) = [R(Tamb) * (1 + α * Tref)] / [1 + α * Tamb]
- Metal material, whose temperature coefficient can be entered by the user (α)
Temperature coefficient beyond operating range <10% accuracy/°C (from 0 to18°C and from 28 to 50°C)
Alarms 2 programmable thresholds with visuel and sound signal

General specifications

Size L x W x H 270 x 250 x 180 mm
Weight 4 kg
Supply 100 to 240 V (50 / 60 Hz)
Batteries (option) Type: Ni/Mh 8.5 Ah
Charging time: 5 h
Battery life: 5,000 measurements at 10 A
Communication ports RS 232
Storage capacity 1,000 measurements identified by numbers
Memory reading directly on the display, via software or printer

Environmental specifications

Reference range 23°C ± 5°C (RH: 45 to 75 % w/o condensing)
Operating reference range 0 to 50°C (RH: 20 to 75 % w/o condensing)
Limit operating range -10°C to +55°C (RH: 10 to 80 % w/o condensing)
Storage temperature limits -40°C to +60°C
IP protection IP53 according to EN60529

Safety specifications

Protections
  • Electronic protection up to 250 V for ‘voltage’ wires
  • Fuse protection for ‘current’ wires
  • Protection against ‘current’ circuit breaking during inductive resistance measurements
Class In accordance with EN 61010-1
Category II, pollution 2
Rated voltage 60 V
Chocks and vibrations EN61010-1
EMC conformity Immunity
  • EN 61000-4-2
  • EN 61000-4-3
  • EN 61000-4-5
  • EN 61000-4-6
  • EN 61000-4-11

EN 61000-4-4
Conducted and radiated emissions:

  • EN 55022, class B
  • EN 61000-3-2
  • EN 61000-3-3