Performances and technical specifications @23°C ±5°C
Uncertainty is given in % of reading + fixed value.
Resistance measurement
Measurement range |
Resolution |
Accuracy / 1 year (23°C ±5°C) |
Measuring current |
Voltage drop |
5 mΩ |
0.1 μΩ |
0.05 % + 0.5 μΩ |
10 A |
50 mV |
25 mΩ |
1 μΩ |
0.05 % + 3 μΩ |
10 A |
250 mV |
250 mΩ |
10 μΩ |
0.05 % + 30 μΩ |
10 A |
2.5 V |
2500 mΩ |
0.1 mΩ |
0.05 % + 0.3 mΩ |
1 A |
2.5 V |
25 Ω |
1 mΩ |
0.05 % + 3 mΩ |
100 mA |
2.5 V |
250 Ω |
10 mΩ |
0.05 % + 30 mΩ |
10 mA |
2.5 V |
2500 Ω |
100 mΩ |
0.05 % + 300 mΩ |
1 mA |
2.5 V |
Automatic or manual selection of measurement range Possible excess over the nominal range:
- 5 mΩ range: + 20 %
- 25 mΩ range: + 20 %
Maximum voltage between the terminals in an open circuit: 7 V Current waveform: Continuous or pulse DC current
Ambient temperature measurement for Tref compensation
Type |
Résolution |
Précision sur 1 an (23°C ±5°C) |
Remarque |
Pt100 |
0,1°C |
0,5°C |
Mesure par Pt100 externe ou valeur saisie au clavier |
Further features
Resistance types |
- Inductive resistances: Coils, transformers, motor windings, twisted cables...
- Non-inductive resistances: Earth bonding, coating, contact résistances, non-twisted cables...
|
Measurement trigger conditions |
Manual or automatic trigger allowing a single operator to be able to perform measurements |
EMFs |
Automatic compensation of EMF parasites before each measurement for a greater accuracy |
Temperature compensation |
- Ambient temperature Tamb, measured with external Pt100 (or entered by the user - Programmed reference temperature Tref, to which the measured value is converted: R(Tref) = [R(Tamb) * (1 + α * Tref)] / [1 + α * Tamb] - Metal material, whose temperature coefficient can be entered by the user (α) |
Temperature coefficient beyond operating range |
<10% accuracy/°C (from 0 to18°C and from 28 to 50°C) |
Alarms |
2 programmable thresholds with visuel and sound signal |
General specifications
Size L x W x H |
270 x 250 x 180 mm |
Weight |
4 kg |
Supply |
100 to 240 V (50 / 60 Hz) |
Batteries (option) |
Type: Ni/Mh 8.5 Ah Charging time: 5 h Battery life: 5,000 measurements at 10 A |
Communication ports |
RS 232 |
Storage capacity |
1,000 measurements identified by numbers Memory reading directly on the display, via software or printer |
Environmental specifications
Reference range |
23°C ± 5°C (RH: 45 to 75 % w/o condensing) |
Operating reference range |
0 to 50°C (RH: 20 to 75 % w/o condensing) |
Limit operating range |
-10°C to +55°C (RH: 10 to 80 % w/o condensing) |
Storage temperature limits |
-40°C to +60°C |
IP protection |
IP53 according to EN60529 |
Safety specifications
Protections |
- Electronic protection up to 250 V for ‘voltage’ wires
- Fuse protection for ‘current’ wires
- Protection against ‘current’ circuit breaking during inductive resistance measurements
|
Class |
In accordance with EN 61010-1 Category II, pollution 2 |
Rated voltage |
60 V |
Chocks and vibrations |
EN61010-1 |
EMC conformity |
Immunity
- EN 61000-4-2
- EN 61000-4-3
- EN 61000-4-5
- EN 61000-4-6
- EN 61000-4-11
EN 61000-4-4 Conducted and radiated emissions:
- EN 55022, class B
- EN 61000-3-2
- EN 61000-3-3
|
|